1D Flatness AveΒΆ
The average value dose parameter is the normalised average between the maximum and minimum profile values taken over the in field area:
\[\cfrac {100} {cax} \cdot \cfrac {max + min} {2}\]
Where max and min are the profile maximum and minimum respectively and cax is the profile centre value.
Protocol invocation name: 1D Flatness Ave
The max, min and cax may be affected by normalisation or windowing.