1D Flatness AveΒΆ

The average value dose parameter is the normalised average between the maximum and minimum profile values taken over the in field area:

\[\cfrac {100} {cax} \cdot \cfrac {max + min} {2}\]

Where max and min are the profile maximum and minimum respectively and cax is the profile centre value.

Protocol invocation name: 1D Flatness Ave